Home Transport
Incubators
Continuous Bedside
Blood Gas Monitoring
Surgical
Gloves
Voyager/LW Transport Incubator

Voyager
Summary
Using the popular Model 20H/20L incubators as a starting point, International Biomedical has lightened the designs even more for the Voyager/Voyager LightWeight. In addition, the new incubator models include a new display, improved thermal function, and further expanded the options for monitoring and ventilation while retaining the same footprint from the 20H/20L that is already compatible with most systems for ground and air transport.

View Product Brochure (PDF)




Specifications (subject to change, dimensions and weights are approximate)
Height
in. (cm)
Width
in. (cm)
Depth
in. (cm)
Weight
Voyager

lbs. (kg)
Weight
Voyager LW

lbs. (kg)
Incubator (W/O Infant Chamber) 10 (25.4)37.5 (95.2)19.2 (48.8)82 (45.8)55.3 (31.5)
Trapezoidal Infant Chamber 11 (28.0)29 (73.7)17 (43.2)17 (7.7)13 (5.9)
Rectangular Infant Chamber 11 (28.0)29 (73.7)17 (43.2)20 (9.1)17 (7.7)
Low-Profile Accessory Module 8 (20.3)31 (78.7)19.2 (48.8)18 (8.2)18 (8.2)
Accessory Module 9.5 (24.1)31 (78.7)19.2 (48.8)20 (9.1)20 (9.1)
Air/Oxy Transport Cart 14 (35.6)38 (96.5)23 (58.4)26 (11.8)26 (11.8)

Mattress Tray Dimensions11.75 x 23.75 x 1 in. (29.8 x 60.3 x 2.5 cm)
Infant Chamber Vertical Clearance9 in. (22.9 cm)
Front Access Door Size8 x 20.25 in. (20.3 x 51.4 cm)
AC Power115V AC, 50-400 Hz, 6 Amps
230V AC, 50-400 Hz, 4.5 Amps
(In both cases, 3 Amps are reserved for accessories.)
External DC Power12V DC, 10 Amps
Internal Battery VoyagerOne 12-Volt 26-Amp-Hour Sealed Lead/Acid Type
Internal Battery Voyager LWOne 12-Volt 13-Amp-Hour Sealed Lead/Acid Type
Battery LifeApproximately 200 cycles
Nominal Battery Recharge (90%) Time Voyager8 Hours with AC Power Connected and Incubator Power Off
Nominal Battery Recharge (90%) Time Voyager LW6 Hours with AC Power Connected and Incubator Power Off
Nominal Battery Operation Time Voyager3 Hours (Chamber 37°C, Ambient 20°C)
Nominal Battery Operation Time Voyager LW1.5 Hours (Chamber 37°C, Ambient 20°C)
Examination Light5 Watt
Temperature Control17.0°C to 38.9°C range, 0.1°C resolution
Digital Display Resolution0.1°C
Digital Display Accuracy±1°C for range of 10.0°C to 45.0°C
Warmup Time9 minutes ±20%, Standard chamber
12 minutes ±20%, Large chamber
Carbon Dioxide ConcentrationLess than 0.5%
Transport and Storage LimitsTemperature: -15°C to 40°C
Humidity: 10% to 95%, non-condensing
Pressure: 50 KPa to 106 KPa
Design StandardsFully compliant with the requirements of the International Standard, IEC 601-2-20 (First Edition 1990) + Amendment 1 (1996).
 |  Parts and Support  |  Contact Us  | 